JRI Maxant has developed a wireless traceability system which is adaptable and user friendly. Its aim is to
maximise security and gain time. This global traceability system SPY RF ® meets perfectly the requirements
of current regulations (HACCP, COFRAC and EU equivalents, ISO, GLP, GMP…) and quality control.
Today, JRI Maxant provides all the components of an efficient traceability system : on site radio
communication survey, metrology, radio communication, hot line, software and maintenance contracts.
Easy to use and delivering reliable results, the SPY RF ® system also offers automatic monitoring of your
laboratory or production plant.
Some of our french users...
FOOD
Kitchen
> St André Les Lille Central kitchen (59)
St André Les Lille Central
Kitchen
Application : control of ambient
temperatures and products in cold
chambers, storage and freezing areas.
> St Malo Police school (35)
> Louvres Hotels (Le Ritz, Ambassador,
> Hôtel du Louvre, Concorde Montparnasse,
> Concorde St Lazare, Lutetia, Concorde Lafayette)
> Charles Perrens Hospital (33)
> Crous of Rennes (Student service) (35)
> Mc Donalds (94)
> Total service stations (multi-site)
> HRC Restaurant " l’Arche " (Elior Group 61)
> Européenne de confiserie (59)
> Holder Moulin Bleu (59)
Supermarkets/dealers
> Total Argedis (Chanteraine site)
> Locaservice (59)
> Leclerc Supermarkets (69)
Industries
Total Argedis
Application : constant monitoring
of chilled cabinets, cold chambers
and ice boxes.
> Delpierre Mer et Tradition (62)
>
>
>
>
>
Milhaud Salting(81)
Berthillon Ice Cream(75)
Condat Paper mill (24)
CBV delicatessen (61)
…
Delpierre
Mer et Tradition
Application : automatic
data archiving in
Industrial processes.
MUSEUMS ARCHIVES
> "Archives Départementales" du Finistère (29)
> Carnavalet Museum of Nice (06)
> Paris National School of Beaux-Arts (75)
> Marc Chagall National Museum (06)
> Museum of Vire (14)
> Fernand Léger Museum (06)
> ...
Marc Chagall National Museum
Application : Temperature and humidity record
for preservation of artwork.
2
…and hundreds of working SPY RF® systems
all over the world.
Ask us for your local distributor details.
TABLE
QUALITY
CONTROL
HEALTH
Blood Transfusion Centers and Blood Banks
> EFS P.A.C.A.
> (Marseille, Avignon, Gap, Aix, Cannes, Bastia, Ajaccio,
> Toulon, Arles)
> EFS NORMANDIE (76)
> (Caen, Rouen, Lisieux, St Lô, Le Havre, Alençon,
> Dieppe, Evreux)
> EFS Pyrénées Méditerranée
Laboratories
> Lexobio Lisieux (14)
Hospitals
> CHU of Brest (29)
> CHRU of Lille (59)
>MEASURES
>RECORDS
>COMMUNICATE
>ARCHIVES
>WARNS
>
>
>
>
>
>
>
>
CHU de Grenoble (38)
CHU de Mulhouse (68)
CHU de Reims (51)
CHU de Nîmes (30)
J. Verdier Hospital (93)
"Assistance Publique des Hôpitaux de Paris" (75)
"Hospices Civiles de Lyon" (69)
Hospital of Dourdan (91)
CHRU de Lille
Application : temperature monitoring
In chambers at -200°C, -80°C, -40 °C
and ovens.
Industries
> Avantis Pharma (37)
> Alliance Healthcare (92)
>
>
>
>
Pharmalog (27)
CERP (Multi-Site)
STEF (Multi-Site)
…
(real time alarm)
Alliance Healthcare
Application : monitoring
of temperature sensitive products.
TRANSPORT UNDER MONITORED TEMPERATURE
> Transports Antoine Group (85)
>
>
>
>
Logidis
Geodis Logistics Ile de France (multi-site)
Flash Europe (57)
…
INDUSTRY
Transports Antoine Group
> EDF (77)
> Siemens Automotive (31)
> Virbac (06)
> Procter & Gamble (41)
>…
Application : real time data monitoring
and product traceability.
3
EXAMPLES OF WORKING SPY RF ® SOL
Si r i u s L i t e v e r s i o n w i t h re p e a t e r
Si r i u s Tr a n s p o r t Ve r s i o n
4
UTION SYSTEMS
Si r i u s St o r a g e v e r s i o n v i a Et h e r n e t n e t w o r k
,78
04
87
,40
8
,7
04
EM
MOD
ng
ldi
i
in
Ma
ing
ild
ary
bu
bu
d
con
Se
04
,78
,78
04
,78
04
Consultation
8
4,7
0
EM
MOD
Consultation
LEGEND
SPY RF ®
SPY RF ® ModeM
8
4,7
0
SPY RF ® RelaY
04
,78
SPY RF ® Visu
Ethernet Network
USB Cable
Radio communication
(200 - 300 m)
Radio communication
(800 - 1000 m)
,78
04
DEM
MO
SPY RF ® AlarM
SPY RF ® Ethernet Modem
Message alert
E-mail Alert
Visual and Audible Alert
Telephone Alert
5
Internet
Server
AMBIENT SPY RF® RECORD
Possibility to add
an anti-theft
on the mounting bracket
Splashproof
LED alarm
Push button
Replaceable battery
Pressure
compensation zone
Input cable protection
LCD screen
Internal memory
Record mode
Pause mode
Radio signal level
Threshold
warning
indicator
Measurement value
and unit
Channel number
6
SPY RF ® T
Ambient temperature
recorder
Reference 06165
Key features
> High color for easy location.
> Economic.
> Splashproof.
T E C H N I C A L
Measurement range
Channel number
Measurement accuracy at 23°C
Resolution
Operating conditions
Power supply
Battery life*
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Conformities
Included
SPY RF ® T+
Ambient temperature recorder
with protruding probe
Key features
Reference 06166
> Has a temperature protruding
probe for a faster response
time.
> High color for easy location.
> Splashproof and
contaminationproof.
Conformities
Included
SPY RF ® TH
Ambient temperature and
relative humidity recorder
Reference 06167
Key features
> Has a low inertia during
measurement variations.
> The probe has integrated
protection.
> Replaceable battery.
Did you
know?
JRI MAXANT offers COFRAC
calibration certificates
for its whole range
of recorders.
From -40°C to + 85°C
1
0.5°C from -30°C to +70°C ; 1°C above +70°C
0,1°C
from -40°C to +85°C
Lithium battery
4 years - not replaceable
Internal (temperature PTC)
1 to 90 minutes
10 000 measurements
1km line of sight
868 Mhz (ISM : Industrial Scientific Medical)
25 mW
No
123X69X30 mm (HxWxD)
IP68
Complies with EN 12830 norm and recyclable
according to EU regulation n° 2005 - 229
User manual
T E C H N I C A L
Measurement range
Channel number
Measurement accuracy at 23°C
Resolution
Operating conditions
Power supply
Battery life*
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Measurement accuracy at 23°C
Résolution
Operating conditions
Power supply
Battery life*
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Conformities
Included
* Depending on operating conditions.
7
F E A T U R E S
from -30°C to + 70°C
1
0.5°C from -30°C to +70°C ; 1°C above +70°C
0,1°C
from -30°C to +70°C
Lithium battery
4 years - not replaceable
Protruding (temperature PTC)
30 seconds to 90 minutes
10 000 measurements
1km line of sight
868 Mhz ISM (Industrial Scientific Medical)
25 mW
Yes
123X69X30 mm (HxWxD)
IP68
Complies with EN 12830 norm and recyclable
according to EU regulation n° 2005 - 229
User manual
T E C H N I C A L
Measurement range
Channel number
F E A T U R E S
F E A T U R E S
From -30°C to + 70°C; from 0 to 100% HR non condensing
1 or 2
0.5°C from -30°C to +70°C and 2% RH
from 10% to 90% RH ; 3,5% above 90%
0,1°C
From -30°C to +70°C
Lithium battery
2 years - replaceable
Internal (temperature et relative humidity)
1 second to 90 minutes
20 000 measurements (10000 per channel)
1km line of sight
868 Mhz ISM (Industrial Scientific Medical)
25 mW
Yes
123X69X35 mm (HxWxD)
IP34
Recyclable according to EU regulation n° 2005 - 229
User manual, wall-mounting bracket
AMBIENT SPY RF ® RECORDERS
ERS
DISTANT SPY RF
SPY RF ® U
Universal recorder
Key features
Reference :
U1 : 06168
U2 : 06169
> Compatible with most
probes and sensors : PT100,
PT1000, 4-20mA, 0-1 V,
contact.
> Monitors all types of
temperature-controlled
equipment
(-200°C to +400°C)
> Allows recording to start
when specific events occur.
> Possibility to have different
types of input in
2-channel model (analogue
or electrical).
T E C H N I C A L
Measurement range
Channel number
Measurement accuracy at 23°C
Resolution
Operating conditions
Power supply
Battery life*:
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Conformities
Included
T E C H N I C A L
Digital recorder
(T° / RH / contact)
Measurement range
> Probes are interchangeable
for easy maintenance.
> Suitable for dry contact
inputs.
> Allows recording to start
when a specific event occurs.
> Possibility to have different
type of input on the
2 channels model
(digital or electrical).
Channel number
Measurement accuracy at 23°C
Resolution
Operating conditions
Power supply
Battery life*
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Conformities
Included
SPY RF ® Health
Dedicated range for pharmacy
and medical biology
developed in collaboration
with the experts of Paris
Hospitals Trust (" AP-HP ")
from SPY RF ® U and N.
Key features
Reference :
U1 : 08634
U2 : 08635
N1 : 08636
N2 : 08637
> Possibility to monitor the hot
and cold spots of your
chamber (using the 2-channel
SPY RF ® Health recorder).
> High accuracy in the range
+2 to +8°C.
> Supplied with COFRAC
calibration certificates
(equivalent to EU standards).
RECORD
F E A T U R E S
from -200°C to + 400°C/4-20mA/0-1V/dry contact
1 or 2
0.3°C (on short MR** only)
0,1°C
From -30°C to +70°C
Lithium battery
2 years - replaceable
External (PT100/PT1000/4-20mA/ 0-20Ma/0-1V/dry contact)
1 second to 90 minutes
10 000 measurements per channel
1km line of sight
868 Mhz ISM (Industrial Scientific Medical)
25 mW
Yes
123X69X35 mm (HxWxD)
IP65
Complies with EN 12830 norm and recyclable according
to EU regulation n° 2005 - 229
User manual, wall-mounting bracket and cable input protection
SPY RF ® N
Key features
Reference :
N1 : 06428
N2 : 06429
®
F E A T U R E S
T probe : from -55°C to +80°C and TH probe :
from -30°C to +70°C from o to 100% HR non condensing
1 or 2
0.5°C (within a range of 50°C) RH : 2% RH from
0 to 90% RH and 3% from 90% to 100% RH (with probe)
0,1°C
From -30°C to +70°C
Lithium battery
2 years - replaceable
External (temperature, relative humidity and dry contact)
1 second to 90 minutes
10 000 measurements per channel
1 Km line of sight
868 Mhz ISM (Industrial Scientific Medical)
25 mW
Yes
123X69X35 mm (HxWxD)
IP65
Complies with EN 12830 norm and recyclable according
to EU regulation n° 2005 - 229
User manual, wall-mounting bracket and cable input protection
T E C H N I C A L
F E A T U R E S
SPY RF ® Health N
SPY RF ® Health U
Measurement range
T : from -55°C to +80°C
T : from -50°C to +180°C
TH : from -30°C to +70°C
Measurement accuracy between
0.3°C and 2% RH
0.3°C with probe
with probe
+2°C and +8°C
Channel number
1 or 2
Resolution
0,1°C
Operating conditions
From -30°C to +70°C
Power supply
Lithium battery
Battery life*
2 years - replaceable
Type of input
External (temperature, relative humidity and dry contact)
Recording interval
1 second to 90 minutes
Internal memory
10 000 measurements per channel
Radio range
1km line of sight
Radio band
868 Mhz ISM (Industrial Scientific Medical)
Power
25 mW
Digital display
Yes
Dimensions
123X69X35 mm (HxWxD)
Protection index
IP65
Complies with EN 12830 norm and recyclable according
Conformities
to EU regulation n°2005 - 229
User manual, wall-mounting bracket, cable input protection,
Included
3 points COFRAC calibration certificate (-18, +2, +8°C),
temperature probe supplied with 3m cable
8
SPY RF ® TC
Temperature recorder for
Thermocouple K probes
Key features
Reference :
TC1 : 06170
TC2 : 06171
> Compatible with all types
of thermocouple K probes.
> Compatible with Hygitherm
probe, specially designed
for pipes.
SPY RF ® C
Recorder for countable events
Key features
Reference :
C1 : 06426
C2 : 06427
> Compatible with most pulse
sensors.
SPY RF ® ReferencE
Reference Device
Key features
> Resolution of 0,01°C.
> Supplied with a 1/3 DIN
PT100 probe.
> High accuracy.
Reference 09509
T E C H N I C A L
Measurement range
Channel number
Measurement accuracy at 23°C
Resolution
Operating conditions
Power supply
Battery life*
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Conformities
Included
T E C H N I C A L
Measurement range
Channel number
Measurement accuracy at 23°C
Resolution
Operating conditions
Power supply
Battery life*
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Conformities
Included
* Depending on operating conditions.
** Measurement range.
9
F E A T U R E S
From 0 to 200Hz
1 or 2
1
FRom -30°C to +70°C
Lithium battery
2 years - replaceable
External (pulses)
1 second to 90 minutes
10 000 measurements per channel
1km line of sight
868 Mhz ISM (Industrial Scientific Medical)
25 mW
Yes
123X69X35 mm (HxWxD)
IP65
Recyclable according to EU regulation n° 2005 - 229
User manual, wall-mounting bracket and cable input protection
T E C H N I C A L
Measurement range
Channel number
Measurement accuracy at 23°C
Resolution
Operating conditions
Power supply
Battery life*
Type of input
Recording interval
Internal memory
Radio range
Radio band
Power
Digital display
Dimensions
Protection index
Conformities
Included
F E A T U R E S
Depending on the probe from +200°C to +1370°C
1 or 2
1°C (for short MR only)
0,1°C
From -30°C to +70°C
Lithium battery
2 years - replaceable
External (thermocouple)
1 second to 90 minutes
10 000 measurements per channel
1km line of sight
868 Mhz ISM (Industrial Scientific Medical)
25 mW
Yes
123X69X35 mm (HxWxD)
IP65
Recyclable according to EU regulation n° 2005 - 229
User manual, wall-mounting bracket and cable input protection
F E A T U R E S
-80°C to +40°C depending on sensors
1
0.1°C from -20 to +40°C and -0.2°C from -80 to -20°C
0,01°C
From -30°C to +70°C
Lithium battery
1 years - replaceable
External (PT100)
1 second to 90 minutes
10 000 measurements per channel
1km line of sight
868 Mhz ISM (Industrial Scientific Medical)
25 mW
Yes
123X69X35 mm (HxWxD)
IP65
Recyclable according to EU regulation n° 2005 - 229
User manual, wall-mounting bracket, cable input protection and PT100 probe
DISTANT SPY RF ® RECORDERS
ERS
ACCESSORIES
R A D I O
AND
C O M M U N I C A T I O N
PERIP
DEVICES
SPY RF ® ModeM
References :
USB : 06337
RS232 : 06214
References :
VDC : 08913
POE : 08917
Connects to your PC, enables communication
with SPY RF ® recorders.
> Communicates with an unlimited number of recorders.
> USB connector.
> Remote alarm driven by integrated dry contact output
(vocal telephonic transmitter, buzzer, light).
> Software or PC failure alarm.
> Can be used with or without SPY RF ® RelaY.
> Supplied with a wall-mounting bracket.
> Operating conditions from 0°C to +70°C.
ModeM Ethernet POE / VDC*
Enables Ethernet connection to maximise distances between
recorders and PC server.
> Operating conditions from 0°C to +40°C.
* POE : Power Over Ethernet
* VDC : Main power supply required (9VDC)
SPY RF® RelaY
Reference 06215
Enables direct RF communication to maximise distances
between recorders and PC server.
> Can relay one or more SPY RF ®.
> Works with external power supply and supplied with a backup battery.
> Include two power ranges : 25 mW to communicate with recorders and
500 mW to communicate with ModeM.
> Supplied with a wall-mounting bracket.
> Operating conditions from -10°C to +70°C
A C C E S S O R I E S
Rubber protector
Reference 06220
Did you
know ?
Wall mounting bracket
(x5)
Reference 06222
Replacement battery pack (batteries + seals) x10
Reference 06569
10
JRI Maxant has
developed a device
to enable on site radio
communication survey,
the SPY RF ® TesT.
A L A R M
D E V I C E S
SPY RF ® AlarM
Reference 06216
Real time warning if a problem occurs.
> Local warning.
> Integrated sound and visual alarm.
> Works with an external power supply and supplied with a backup
battery.
> Pushbutton for local alarm acknowledgement.
> Relay output to trigger remote alarm
(vocal telephonic transmitter …)
> Warns SIRIUS software when the battery is low.
> Supplied with a wall-mounting bracket.
> Operating conditions from -10°C to +70°C.
Audible and visual alarm device
Reference 06570
>
>
>
>
Warns if a problem occurs.
Works with an external power supply.
Connects directly to SPY RF ® ModeM or to SPY RF ® AlarM.
Audible and visual Alarm .
Vocal Telephonic Transmitter (VTT)
Reference 03833
The vocal telephonic transmitter is an ideal component for any
alarm installation. In vocal mode it will call from 1 to 4 phone
numbers and inform them clearly on the nature of the incident.
Thanks to its vocal help, the VTT is easy to use.
> Need a digital telephone line.
> Customization of identification, alarms and acknowledgement
of alarms.
> Need external power (ref. 04729).
Reference 00998
GSM ModeM
The GSM ModeM is used with SIRIUS Storage,
SIRIUS Transport, and enables SMS messages.
SPY RF ® Visu
Reference 06216
Visualisation of measurements and alarms,
(can be used only with SIRIUS Transport).
> Real time warning if a problem occurs.
> Works with an external power supply and supplied
with backup battery.
> Sound and visual alarm.
> Rolling measurement display.
> Can manage up to 20 SPY RF ® recorders.
11
ACCESSORIES AND PERIPHERALS SPY RF ®
HERALS SPY RF®
ACCESSORIES
P
R
O
B
E
S
A
N
D
AND
S
E
N
S
O
PERIP
R
S
Technological features of probes
The JRI Maxant Metrology service recommends the use of two types of probes in order to ensure accurate
measurements. The PT100 probe is recognised as the industrial standard. Available in a range of accuracy levels
(B, A, 1/3 Din, 1/10 Din).
The digital probe stores its gauging coefficient. It makes it interchangeable without any gauging or calibration
constraints on site. Indeed digital probes are gauged in optimal conditions within our laboratory.
All our probes have their binder connector to be connected to the SPY RF ® Recorders.
PT100 Probes
Reference 08513 Stainless steel CL.B ( 0,30°C to 0°C), 3 wires, ø5x40 mm, 5cm silicone cable (-50+180°c).
Reference 00966 Stainless steel CL.B ( 0,30°C to 0°C), 3 wires, ø5x40 mm, 3m silicone cable (-50+180°c).
Reference 08512 Stainless steel CL.B ( 0,30°C to 0°C), 3 wires, ø5x40 mm, 15m silicone cable (-50+180°c).
PT100 Piercing probe
Reference 00967 CL.B ( 0,30°C to 0°C), 3 wires, 3m teflon silicone cable, case ø4x150 mm,
Stainless steel handle ø9x100 mm (-50+180 °c).
PT100 Probe (-80°C)
Reference 00968 Stainless steel ø6x50 mm, CL.B ( 0,30°C to 0°C), 3 wires, 3m teflon/teflon cable (-90+160°c).
For SPY RF ® U
PT100 Cryogenic probe
Reference 06621 CL.B ( 0,30°C to 0°C), 3 wires, ø3x18 mm, 2m teflon/teflon cable (-250+150°c).
PT100 High temperature probe
Reference 08891 Stainless steel ø6X50 mm, 3m cable, 3 wires, interwoven stainless steel & fibreglass (0+400°c).
PT100 1/3 DIN probe
Reference 08511 ( 0,1°C to 0°C), 3 wires, ø3x50mm, 3m cable (-90+250°C).
PT100 Flat cable probe
Reference 09520 CL.B ( 0,30°C to 0°C), 3m cable, 3 wires, ø5x15 mm (-50+125°C).
Digital probes
Reference 06556 (-40+80°C) 3m flat cable, from 0,5°C on the whole range to 0,2°C on a smaller range.
Reference 08872 (-40+80°C) 50cm flat cable with IP67connector, from 0,5°C on the whole range
to 0,2°C on a smaller range.
For SPY RF ® N
TH Digital probes
Reference 09471 (-40+80°C / 0-100% RH*) extension lead available (see page 13).
Reference 09501 With protecting coating** (-40+80°C / 0-100% RH*) extension lead available
(see page 13).
* No condensing.
** Saline or corrosive environment.
Did
you know ?
For measurement channels composed
of SPY RF ® U recorders associated
with PT100 or digital probes,
our metrology laboratory guarantees
the following Maximum Permissive Error :
For use across a 35°C temperature
range between the limits of
-20 and +40°C : 0.3°C
For use across a 5°C temperature
range between the limits of
-20 and +40°C : 0.2°C
12
TH Transmitter
Reference 05886 Wall mounting 2x4-20mA (0 +50°c/0-100% RH*),
without power supply, or universal adapter.
RH Transmitter
Reference 05887 Wall mounting 4-20mA (0-100% HR*), without power supply,
or universal adapter.
CO2 Transmitter
Reference 08991 (0 to 20%), 0.3% + 2% of the reading.
For SPY RF ® U
Supplied with 1power supply and a 2m cable for SPY RF ® U.
Mounting bracket for CO2 Transmitter
Reference 08337
∆P Transmitter
Reference 08371 Measurement range from o to 100 Pa, ABS shell, IP65, 2 visual
and sound alarms, requires power of 230 V, and 2 inputs (4-20 mA
or 0-10 V), universal adapter to be ordered seperatly.
Hygitherm TCK probe
Reference 06499 Specially developed sensor to monitor the temperature
of sanitary water in pipes (-20+75°C).
For SPY RF ® TC
TCK Probe
Reference 08537 ø6x400 mm (-200+1100°C), inconel duct, insulated soldering.
Power Supply Failure Detector (DCS)
Reference 08784 Alerts to mains power supply failure. This sensor links the mains
For SPY RF ® U ou N
power supply to the SPY RF ® recorder.
When a power supply failure occurs, the DCS sends the information
to SIRIUS software which triggers an alarm.
P
R
O
B
E
S
A
C
C
E
S
S
O
R
3 m extension lead
Universal adaptator
Reference 08873 For digital T probe with IP67 connector
Reference 00949 (5m) with connector IP67
for SPY RF ® N.
Reference 06465 For digital TH probe with IP67 connector
for SPY RF ® TH.
I
E
for SPY RF ® N or U.
12VDC Power supply
Reference 04729 For transmitters,
up to 4 transmitters 4/20 mA or 1 Vocal
Telephone Transmitter.
12 m extension lead
Reference 08874 For digital T probe with IP67 connector
for SPY RF ® N.
Référence 09468 For digital TH probe with IP67 connector
for SPY RF ® TH.
24VDC Power supply
Reference 04732 For transmitters,
up to 16 transmitters 4/20 mA.
13
S
ACCESSORIES AND PERIPHERALS SPY RF ®
HERALS SPY RF®
R A N G E
O F
S I R I U S
JRI Maxant has developed a user friendly
software range which enables configuration
of an unlimited number of recorders, as well
as management of data and alarms.
SIRIUS range is available in 3 versions to
meet your specific needs.
SIRIUS Lite,
simple mono-user software
The benefits of SIRIUS Lite :
>
>
>
>
>
>
>
Management of several SPY RF ® recorders.
Collects recorded data From SPY RF ® recorders.
Data are secured (not falsifiable).
Displays graphs and tables together with statistics
Data can be exported to Excel.
Does not require a dedicated PC.
Manages alarms via either a dry contact or a SPY RF ®
AlarM device (telephone call, flashing light, buzzer).
14
Did
you know ?
This system enables
data access via Internet,
helping you to react
quickly should
a problem occur.
SIRIUS Storage,
advanced software dedicated to storage applications
The benefits of SIRIUS Storage compared to Sirius Lite :
> Quick overview of the system via a site mapping function.
> Enables remote alarms (SMS, e-mail, telephone call, light,
buzzer …).
> Programmable task planner: alarm inhibition, upload of data ,
audit trail and alarm report printing, defrost cycles …
> Choice of 3 accessibility levels: administrator, batch manager
and user.
> Can accept several SPY RF ® ModeM devices.
> Capable of monitoring large and distant systems.
> Available in mono-user and multi-user versions.
> Complies with FDA CFR 21 part 11.
15
RANGE OF SIRIUS SOFTWARE
S O F T W A R E
R A N G E
O F
S I R I U S
SIRIUS Transport,
designed for single or multi step transport applications
The benefits of SIRIUS Transport compared
to Sirius Lite :
> Enables access to information while SPY RF ® recorders remain
in vehicle.
> Enables configuration of transport data sheet
(type of transport, details of sender and receiver, vehicle
registration, driver name, stages, dates).
> Allows tracing of each delivery step on the graph.
> Choice of 2 accessibility levels : administrator and user.
> Can accept several SPY RF ® ModeM devices.
> Graph annotation function following alarm activation.
> Local alarm system.
> Allows management of data archives.
> Available in mono-user and multi-user versions.
Did
you know ?
JRI Maxant offers
both calibration
software for SPY RF devices ®
and temperature mapping
software complying
with the
NFX 15140 norm.
16
Technical features of SIRIUS software
Sirius Lite
Sirius Stockage
Sirius Transport
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
06450
Mono-user : 06448
Multi-user : 06449
Mono-user : 06451
Multi-user : 06452
Recorder management
Reading last measurement
Upload memory of one or all SPY RF ® recorders
Management of several modems
Configuration
Management from several PCs
Software control panel with site map
Software control panel without site map
User and batch managment (CFR 21 Part 11)
Upload measurements
Reading of the last measurement
Inhibition d'alarme
Automatic printing of reports
Defrost cycle management
Alarm acknowledgement function including cause of alarm
Typing a cause of alarm
SMS
E-mail
Telephone calls
Other (via dry contact)
Audit trail report
Configuration and recording
Archive management
User friendly data transfer (E-mail)
Measurements
Select the channel to be displayed
Zoom and statistics
Event display
Secondary scale
Super-imposed graphs
Cursor display
Export pictures to Word
Save an archive
Values and statistics
Alarm reports
Export to Excel
Reference
Reference 06632
√
√
√
√
√
√
√
√
√
√
√
SIRIUS Point,
Enables gauging of SPY RF ® recorders (except SPY RF ® N
and SPY RF ® C) and printing of gauging certificates.
17
RANGE OF SIRIUS SOFTWARE
S O F T W A R E
SPY
RF®
WIRELESS
SYSTEMS
JRI MAXANT SERVICES
< Installation
In France and worldwide, various services are offered : on site
radio communication survey, installation, commissioning,
IQ/OQ qualification of systems, on site metrology and
training.
< Metrology
Our metrology laboratory provides : COFRAC-linked
checking reports, gauging certificates, temperature
mapping reports, as well as calibration certificates for
devices of any brand.(Temperature, relative humidity,
CO2, pressure…)
< Training
JRI Maxant offers different levels of training : technical
maintenance and metrology, installation and commissioning,
use and administration of a wireless system.
The JRI Maxant training can be integrated in your annual
training/education plan.
< Maintenance contract available through your local
distributor
●
●
●
18
Preventive maintenance.
Corrective maintenance.
Metrology.
F A Q
1
A S T H E D EV I C E U S E S R A D I O F R E QU E N C Y, W I L L I T A F F E C T
S E N S I T I V E E L E C T R I C A L E QU I P M E N T ?
NO : If your electrical equipment is conform to the European norms then the
radio frequency waves will not interfere with it.
2
3
A R E R A D I O F R E QU E N C Y D EV I C E S D A N G E RO U S F O R H U M A N S ?
NO : The energy emission of radio frequency devices is very low. It is around
40 or 50 times less than that emitted by a mobile phone.
4
5
I S T H E R E A L I M I T TO R A D I O WAV E T R A N S M I S S I O N
R A N G E I N S I D E BU I L D I N G S ?
NO : Modern Radio Frequency technology enables transmission
up to 5 km line of sight. JRI Maxant offers technical solutions
to overcome physical obstacles and long distances.
IS IT POSSIBLE TO INSTALLARADIO FREQUENCY SYSTEM IN ANY
LOCATION, WITH ANY TYPE OF WALLS ?
YES : We can assure you that radio frequency works everywhere
with any type of walls even if they are thick, providing that the installation
recommendations are followed.
6
7
ARE RADIOFREQUENCY DEVICES SUITABLE FOR A MEDICAL
ENVIRONMENT ?
YES : Radio waves function within the ISM 868 MHz bands
(Industrial Scientific and Medical).
DOES THE USE OF THESE DEVICES
REQUIRE AUTHORISATION ?
NO : You are allowed to use them anywhere because
they comply with the European norm ETS 300-220.
ARE MEASUREMENTS DELIVERED BY RADIOFREQUENCY SYSTEMS
AS ACCURATE AS THOSE FROM AN ELECTRONIC DEVICE ?
YES : Radio Frequency doesn’t interfere with the measurement, therefore
these devices have the same accuracy as equivalent electronic wire devices.
19
MARQUE JAUNE Creation - www.marque-jaune.com - June 2008.
JRI Maxant
Rue de la Voivre
Parc d’Activités Technoland - BP 51027
25490 FESCHES LE CHÂTEL CEDEX - FRANCE
Ph. : +33(0)3 81 30 68 04
Fax : +33(0)3 81 30 60 99
www.jri.fr - E-mail : [email protected]
JRI Maxant - Head office : 116, quai de Bezons - BP 20085 - 95101 ARGENTEUIL CEDEX - FRANCE
S.A.S. 4 000 000 € CAPITAL - R.C.S. PONTOISE 380 332 858
To m a k e y o u r t ra c e a b i l i t y, o u r p r i o r i t y…

Benzer belgeler