Sisdoz_m20_cl_İşletme Bakım Klavuzu

Transkript

Sisdoz_m20_cl_İşletme Bakım Klavuzu
M20W CL
Instruction Manual
1
2
0.23
ZERO
GAIN
ppm
SET
CAL
FUN
M20 W
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
1
The box contain
Measure instrument, instruction manual, mounting accessories, fuse.
General features
Measure range
Functions
Resolution
Display
Accuracy
Controls
Set-point
Outputs
Hysteresys
Delay time
P.W.M.
Power supply
Consumption
Fuse
Box
Dimensions
Weight
0-10.00 ppm
Measure, control and regulation of free Chlorine concentration
±0.01 ppm
LCD 3 ½ DIGIT
± 1% F.S.
Keyboard (5 keys)
2 points ON/OFF 5A to 230vac
4-20mA selectable for recorder or proportional to the setpoint 2
Adjustable for each setpoint
Adjustable for each setpoint
Selectable on set point 2 in time/pause way
230 Vac 50Hz (optional 110/ 24 Vac)
5W
315 mA delayed
Plastic IP65
180x115x75 mm
560 gr.
Attention
Before to make any operation on the isntruments to verify that the power
supply cable isn’t connected
Check that the electric system voltage correspond with value on the instrument
label
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
2
Dimensions
0.23
1
180 mm
2
ZERO
GAIN
ppm
SET
CAL
FUN
M20 W
115 mm
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
75 mm
3
Electrical connections
Power supply
To connect the power supply wires to the clamps 1 and 3 (2 ground) and tourn-on the
instrument. The display shows the software revision and successfully the instrument go in
measure and is ready to follow the normal functions.
Description
Display
Led relays
status
Led zero
calibration
1
2
0.23
GAIN
ppm
Up – Down
keys
SET
key set-point
menu
Led gain
calibration
ZERO
M20 W
Key calibration
starting
CAL
FUN
key menu
functions
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
4
Calibration
Connect the probe to the instrument and leave to flush water with 0 ppm of chlorine for
at least 5 minuts to stabilize the probe.
 Push CAL key. The red led ZERO flashing and shows the probe zero adjustment.
 Pushing UP and DOWN keys to adjust the display value since to show 0.00 ppm.
Therefore push key CAL to confirm.
 Flush water with ppm value known (DPD1 colorimetric test). The red led GAIN flashing
and shows the probe gain adjustment.
 Pushing UP and DOWN keys to adjust the display value since to show the value of
chlorine concentration. Therefore push key CAL to terminate calibration procedure.

If during calibration you don’t push any key for 60 seconds the instrument automatically
back in measure condition.
Set-Point adjustment
Set-point 1
 Push SET key, the corresponding green led flashing, the display shows the value
previously set (default 1.00 ppm). With UP and DOWN keys adjust the display value since
to obtain the desired value. Push key SET to confirm the value and to pass at the
hysteresys setting.
 The display shows the value previously set (default 0.1 ppm), through Up and DOWN
keys is possible to modify this parameter by steps of 0.1 ppm. Push SET key to confirm
the value and to pass at the delay time setting.
 The display shows the delay time value (seconds) to active or deactive the set-point
(default d00, 0 seconds). Through UP and DOWN keys it’s possible to modify the value
since to 99 seconds. Push SET key to confirm the value and pass to the Set-point 2.
Set point 2
 Finished the set-point 1 setting display shows the set-point 2 value previously set and
the green led 2 flashing. Following the same operation about the set-point 1 setting is
possible to modify the set-point 2 value, hysteresys and delay time. Push SET key to
confirm the value and back to the measure mode.
If during the set operations doesen’t push any keys for at least 60 seconds the instrument
automatically back in measure mode.
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
5
Functions
Through the activation / deactivation of the function is possible to change the statement
of the instrument for a best flexibility.
Active / Deactive instrument functions
 Push FUN key, the display shows F1 with the function state (d = Deactived, A =
Actived).
 Pushing UP or DOWN keys is possible to change the function state.
 To confirm choice push FUN key, if you changed the function state the instruments
memorized the choice and back in the measure mode, otherwise it pass to the succesfully
function, since to the function F8, therfore back in the measure mode.
If during the set operations doesen’t push any keys for at least 60 seconds the instrument
automatically back in measure mode.
Direct or reverse mode
Function
Set-point 1 mode: direct
F1_d
Function
Set point 1 mode: reverse
F1_A
Function
Set-point 2 mode: direct
F2_d
Function
Set point 2 mode: reverse
F2_A.
Normally both the set point are set for a direct mode regulation, therefore the output relay
is actived when the measure is lower then the set-point value. To change this condition is
necessary to active the F1, about the set-point 1, and active the F2 for the set-point 2.
Hysteresys or double threshold
Function
set point 1 with hysteresys
F3_d
Function
set point 1 with double threshold
F3_A
Function
set point 1 with hysteresys
F4_d
Function
set point 1 with double threshold
F4_A
Hysteresys function creates a no-operation range around the set-point value. In this area
the relay doesn’t change the state (ex.: 0.5H creates a renge of 0.5 ppm).
If you active the function F3 (set-point 1) and F4 (set-point 2) both the set-point change
the functional mode in double threshold. In this way you create a window around the setpoint value. Example: set-point set on 1.00 ppm and hysteresys on 0.5H, in this way you
create a window between 0.50 and 1.50 ppm. Into this range the realy is in off, over this
range the relay is in on. Obviously is possible to reverse this condition changing the
parameters F1 and F2. This solution is useful if you need to create an alarm window.
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
6
Function F5 unused
4-20mA current output
Function
Realy on set point 2 ON/OFF activation and 4-20mA output for recorder
F6_d
Function
Relay on set point 2 proportional time/pause and 4-20mA output
F6_A
proportional to the set point 2 value
With F6 deactived (default) the set-point 2 relay work ON/OFF and the 4-20 mA output
signal is proportional to the measure range (4 mA = 0 ppm; 20 mA = 10 ppm).
When F6 is actived the set-point relay work in proportional time/pause way and the 4-20
mA output signal is proportional to the set-point value. In this way the signal is 4 mA
when the measure in on the set-point value. Throgh the F2 function is possible to decide
the “direction” of 4-20 mA signal .
Function
4-20 mA proportional signal on set point 2 with range regulation of 1.00
F7_d
ppm
Function
4-20 mA proportional signal on set point 2 with range regulation of 0.50
F7_A
ppm
With F7 function is possible to decide the range value about the proportional signal on setpoint 2.
Function F8 unused
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
7
Troubleshooting
Malfunction
Possible cause
Solution
Check the electrical
connections
Check if the power supply
No power supply
Display switch off
printed on the instrument
label is correspondent with
local electrical net
Internal fuse is breack
Change the fuse
To be sure, check the
The chemical parameter
parameter with a portable
value is constant
instrument or with a
Measure on the display is fix
colorimetric kit
(doesn’t change)
Repeat the probe calibration
The signal from the probe
and if the problem remain
doesn’t change
change the probe
Check the electrical local
Electrical disturbs from
net. Control the heart
electrical local net
connection.
Measure on the display
Check the instrument,
change continuously
putting the probe into a
(measuring jumps)
Electrical disturbs into the
buffer solution, if the
measuring liquid
instrument measure
correctly eliminate the
electrical disturb.
Buffer solution are defective Change the buffer solution
Isn’t possible to complete
the calibration procedure
Probe is defective
Change the probe
Setpoint value is wrong
Change setpoint value
Change the setpoint
The setpoint relay doesn’t
function mode, direct or
close the contact
Setpoint “direction” is wrong
reverse, from the menu
functions
If isn’t possible to back the instrument in a correct functionality following the prevoius
indications please contact your local distributor.
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
8
EASY SETTING
Set-Point pH
Dosing = Chlorine
Set-point = 1.00 ppm
1.10
1.00
X5
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
9
Calibration
0.05
0 ppm
0.00
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
10
1.05
Example: Chlorine
concentration =
1 ppm
1.00
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
11
Note:
Sisdoz Dozlama Sistemleri Üretim ve Pazarlama Ltd. Şti.
Bostancı Yolu Cad. Keyap Sitesi A-2 Blok No: 12 Y.Dudullu, Ümraniye, İstanbul
Tel: 0216 526 54 70 Faks: 0216 526 54 71 [email protected]
www.sisdoz.com.tr
12

Benzer belgeler